Abstract

Ultrasound (US) is the primary imaging modality for the assessment of transplanted kidneys. This study aims to investigate the ability of conventional US and contrast-enhanced US (CEUS) in assessing renal allograft function and prognosis. A total of 78 consecutive renal allograft recipients were enrolled. Patients were classified as normal allograft function (n=41) and allograft dysfunction (n=37) groups. All patients underwent US and parameters were measured. The independent-samples t-test or Mann-Whitney U test, logistic regression analysis, Kaplan-Meier survival plots, and Cox regression analysis were used. In multivariable analysis, cortical echo intensity (EI) and cortical peak intensity (PI) were determinant US parameters for renal allograft dysfunction (p=.024 and p=.003, respectively). The combination of cortical EI and PI showed an area under the receiver operating characteristic curve (AUROC) of .785 (p<.001). Of 78 patients (median follow-up: 20mo), 16 (20.5%) exhibited composite end points. Cortical PI had a general prediction accuracy with an AUROC of .691, sensitivity of 87.5%, and specificity of 46.8% at the threshold of 22.08dB in predicting prognosis (p=.019). The combination of estimated-glomerular filtration rate (e-GFR) and PI in predicting prognosis showed an AUROC of .845 with a cut-off value of .836, sensitivity of 84.0%, and specificity of 67.3% (p<.001). This study indicates that cortical EI and PI are useful US parameters for evaluating renal allograft function and e-GFR combined with PI may provide a more accurate predictor of survival.