Abstract
In order to assess the functioning of the pivoting disc of the Lillehei-Kaster prosthetic valve, a cinefluoroscopic technique using multiple angulations was developed. The disc was detected in 240 (99%) of the 242 valves studied (126 aortic, 107 mitral, and nine tricuspid valves). The maximal-opening angle of the disc was measured in 75% of the aortic but only 20% of the mitral valve prostheses in the 182 valve studies in which this was attempted. Prosthetic dysfunctions were suspected clinically in three aortic and three tricuspid valves and confirmed with the cinefluoroscopic technique. This cinefluoroscopic approach constitutes a rapid, noninvasive, and sensitive evaluation of the function of the Lillehei-Kaster prosthetic valve in the aortic position.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.