Abstract
Transthoracic Doppler echocardiography examination has become an integral part of the investigations performed in patients with mitral valve prostheses. The limitations of the transthoracic approach are well documented. Transesophageal echocardiography provides a unique window for achieving a clear view of the mitral prosthesis. This study shows the usefulness of transesophageal echocardiography in clinical practice for assessment of patients with a mitral valve prosthesis. This technique demonstrated an abnormality in 48% of patients who had normal results on transthoracic examination. The overall sensitivity of transesophageal echocardiography was 96%. Transesophageal echocardiography constitutes an essential part of a comprehensive two-dimensional/Doppler echocardiographic examination in patients with suspected malfunction of mitral prostheses.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.