The use of magnetic resonance imaging (MRI) in the exploration of the orofacial region of patients who have orthodontic appliances can be disturbed by artifacts, with a theoretical risk of displacement and overheating of the different materials used in odontology.The purpose of this study was to evaluate the thermal effects and the risk of displacement induced by MRI on certain metallic devices used in orthodontics.The results show a very moderate increase in temperature of the materials during MRI exposure, less than 1̊C, in particular in the metal wire linking the brackets. The maximal forces observed were on the order of 0.27 N. The risk of detachment and displacement seems to be nonexistent at 1.5 Tesla when the usual recommendations are respected. Temporary removal of the wire and verification of the adhesive is recommended. However, priority should be given to nonmagnetic or slightly magnetic material to limit artifacts as much as possible when exploring the head region.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.