Abstract

ABSTRACT Many professionals use the handheld dynamometers (HHDs) to assess isometric strength. This study aimed to evaluate the validity, reliability, and sensitivity to change of a DiCI (a new HHD) for the strength measurement of knee and hip muscles and their ratios. Thirty-five recreational athletes completed quadriceps, hamstring, and hip adductors and abductors strength tests in two identical sessions with 1-week interval between them. A fatigue test was carried out for the quadriceps and hamstring. Consistency between DiCI and MicroFet2 was excellent (r > 0.939) for all strength tests. Relative reliability analysis showed good to excellent ICCs (ICC’s isolated muscle range = 0.93 to 0.96 and ICC’s ratios range = 0.70 to 0.82). Sensitivity to change showed large change magnitudes for quadriceps and hamstring fatigue (−1.89 and −1.26, respectively). DiCI is valid and reliable to assess knee and hip strength and sensitive to fatigue change.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.