Abstract

Few studies have reported on the validity of a sensor-based lower-limb kinematics evaluation during the timed up and go (TUG) test. This study aimed to determine the validity of a wearable gait sensor system for measuring lower-limb kinematics during the TUG test. Ten young healthy participants were enrolled, and lower-limb kinematics during the TUG test were assessed using a wearable gait sensor system and a standard optical motion analysis system. The angular velocities of the hip, knee, and ankle joints in sit-to-stand and turn-to-sit phases were significantly correlated between the two motion analysis systems (R = 0.612–0.937). The peak angles and ranges of motion of hip, knee, and ankle joints in the walking-out and walking-in phases were also correlated in both systems (R = 0.528–0.924). These results indicate that the wearable gait sensor system is useful for evaluating lower-limb kinematics not only during gait, but also during the TUG test.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.