Abstract

To estimate validity of two point of care glucometers for the diagnosis of neonatal hypoglycemia and to determine the glucometers cut off values for which standard laboratory confirmatory test are no longer needed. Prospective study. A tertiary care, university hospital in Bangkok, Thailand. The study included 180 blood specimens from 166 high risk neonates aged between 1-24 hours. On average, most of the blood glucose read-outs from the Nova StatStrip and SureStep were higher than laboratory plasma glucose throughout the glucose range with mean differences (SD) of 11.2 (8.4) mg/dL and 13.7 (6.8) mg/dL, respectively. Sensitivity of Nova StatStrip and SureStep were 62% and 53.3%, respectively. Specificity and positive predictive value of both glucometers were 100%. Negative predictive values of both glucometers were approximately 85%. The cut-off levels with 100% negative predictive values were 63 mg/dL and 62 mg/dL for Nova StatStrip and SureStep, respectively. None of the glucometers in this study has sufficient validity to replace laboratory testing in diagnosing hypoglycemia. Confirmatory plasma glucose for diagnosis of hypoglycemia is needed when POC readings are between 39 and 63 mg/dL for Nova StatStrip and between 39 and 62 mg/dL for SureStep.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.