Abstract

To evaluate the effect of treatment on forearm rotation, torque muscle strength can be assessed using an isokinetic device (IKD) or a wrist dynamometer (WD). The aims of this study were 1) to determine concurrent validity and intra- and inter-rater reliability using the WD, and to examine correlations between WD and IKD in different positions; and 2) subsequently, to establish the intermethod reproducibility between WD as a handheld (HHD) or fixed device. We conducted a cross-sectional study in which torque strength was measured in healthy participants by two observers using an IKD and a WD. Study endpoints were concurrent validity (Pearson’s r), intra- and inter-rater reliability, intermethod reproducibility (intraclass correlation coefficient: ICC) and measurement error (limits of agreement: LoA). Concurrent validity ranged, in the 2 studies assessing it, from r 0.37 to 0.52 for pronation and from r 0.50 to 0.82 for supination, with wide 95% confidence intervals. ICC for intra-rater reliability for pronation ranged from 0.85 to 0.91 and for supination from 0.91 to 0.95. ICC for inter-rater reliability for pronation ranged from 0.84 to 0.96 and for supination from 0.92 to 0.96. Despite the excellent intra- and inter-rater reliability and intermethod reproducibility for the WD-HHD and fixed WD, validity was low when compared to IKD and wide LoA indicated a high measurement error of approximately 20%. These results suggest that the WD cannot replace the IKD isometric mode for pronation and supination. Level of evidence2.