Abstract

AimThe aim of this study was to establish the validity of an external measurement system (the Hawk goniometer). This is a digital device which measures joint ranges compared to the universal goniometer for the measurement of shoulder range of motion in healthy adults. Material and methodsa correlational study with 157 healthy volunteers completed six shoulder movements (forward flexion, abduction, extension, adduction and internal and external rotation) with each shoulder. The degree of agreement between each goniometer and Hawk measurement was assessed using Intra-class Correlation Coefficients (ICC) and Bland-Altman 95% limits of agreement (LOA). Resultsthe tests showed a very strong relationship between the readings of both devices (CI between 0.81 and 0.99) and there were no significant differences between the mean readings of both devices. An intraclass correlation coefficient (ICC) of above 0.9 was obtained, indicating a high intra-evaluator reliability of the Hawk goniometer in repeated measurements of shoulder range of motion. Conclusionsthe Hawk goniometer is a valid and reliable element for the objective measurement of the range of motion at the shoulder joint.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.