Abstract

In this paper we present a prototype of a new class of detectors, metallic magnetic calorimeters operating at cryogenic temperatures, which we are developing for absolute activity measurement of low-energy-emitting radionuclides. We explain the detection principle and give a detailed description of the realisation of the prototype, containing an 55Fe source inside the detector absorber. The analysis of first data taken with this detector is presented and the result of activity measurement compared with liquid scintillation counting. Some ways for reducing the uncertainty that can be achieved with this new technique are proposed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.