Abstract

Introduction A low-vision assessment (LVA) is central to developing a vision rehabilitation plan. However, access to LVAs is often limited by the quantity and geographic distribution of low-vision providers, as well as patient-centred transportation challenges. A tablet-based LVA tool kit, delivered virtually, has the potential to overcome many of these barriers. The purpose of this research was to validate a key component of the tablet-based tool kit – a commercially available iPad visual acuity (VA) test (Eye Chart Pro) iPad app – in a low-vision population. Methods Participants with low vision (n = 26) and those who were normally sighted (n = 25) underwent VA testing with both the iPad VA test application and the Early Treatment Diabetic Retinopathy Study (ETDRS) chart. The VA data were compared using a t-test, linear regression and Bland–Altman analysis. Results There was no significant difference in the mean absolute difference in VA (log of minimum angle of resolution (logMAR)=0.11; p = 0.82). Eye Chart Pro and Standard ETDRS Chart measures were also not significantly different (p = 0.98). However, there were significant differences between test methods in the low-vision group and the normally sighted group (p > 0.0001 and p = 0.007, respectively). The Bland–Altman analysis showed a mean bias (difference) of –0.0005 logMAR between methods, and 95% limits of agreement of 0.298 and –0.299 logMAR. Discussion The ETDRS chart function on the Eye Chart Pro application can reliably measure VA across a range, from normally sighted patients to those with low vision.