Abstract

Mobius3D and MobiusFX are model-based verification tools for treatment plan dose calculation and treatment delivery. The software facilitates patient-specific quality assurance by extracting data from linear accelerator treatment log files and performing a 3D dose calculation on the original patient CT dataset using an independent collapsed cone algorithm. In this study, we evaluate the ability of the Mobius system to detect linear accelerator-related errors compared with existing measurement-based systems, namely the ArcCHECK® and 3DVH® systems. Three original treatment plans and 47 plans with introduced delivery errors, for a total of 50 plan deliveries, were investigated. The results from this study demonstrated comparable gamma passing rates and error detectability between the Mobius and ArcCHECK® systems while the 3DVH system generally exhibited a lower sensitivity. This work also demonstrated the ability of the Mobius system to detect delivery errors of down to 2° collimator rotation, 1mm MLC bank offset and 10mm collimator jaw offset.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call