Abstract

The objective of the study was to determine the accuracy of the BP W200-1 device for blood pressure measurement at the wrist developed by the Microlife Company. The BP W200-1 model features a 'Microlife Average Mode', which uses a 'weighted' average of a minimum of three consecutive blood pressure (BP) readings taken 15 s apart. Device evaluation was carried out using the International Protocol of the European Society of Hypertension. Monitor performance was assessed in relation to patients' sex, age, wrist circumference, and systolic (SBP) and diastolic blood pressures (DBP). The BP W200-1 comfortably passed all three phases of the protocol for SBP and DBP. The device performed well in all ranges of SBP and DBP. Mean blood pressure differences for the BP W200-1 monitor were -1.2+/-4.5 mmHg for SBP and -0.2+/-3.1 mmHg for DBP. In multivariable analyses, the SBP and DBP discrepancies between device and observer were unrelated to age, sex, wrist circumference, and blood pressure level. These data show that the Microlife BP W200-1 wrist monitor satisfies the recommended European Society of Hypertension accuracy levels for SBP and DBP and therefore can be recommended for clinical use in the adult population. Its performance is uniform across subgroups of patients with different clinical characteristics.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.