Abstract

The aim of this study was to evaluate the accuracy of the InBody BPBIO250 oscillometric upper-arm professional office blood pressure (BP) monitor in general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018). Subjects were recruited to fulfil the age, gender, BP, and cuff distribution criteria of the AAMI/ESH/ISO Universal Standard in general population using the same arm sequential BP measurement method. Two cuffs of the test device were used for arm circumference 22-32 (medium) and 32-42 cm (large). Ninety-seven subjects were recruited and 88 were analyzed. For validation criterion 1, the mean ± SD of the differences between the test device and reference BP readings was 1.9 ± 5.9/0.9 ± 4.9 mmHg (systolic/diastolic). For criterion 2, the SD of the averaged BP differences between the test device and reference BP per subject was 4.78/4.31 mmHg (systolic/diastolic). The InBody BPBIO250 professional oscillometric device for office BP measurement fulfilled all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) in general population and can be recommended for clinical use.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.