Abstract

The model of the Geiger-Muller counter as the internal part of BDMG-04-02 detection unit in the calibration fa-cility UPGD-2 was developedin MCNP6.2. The different methods are used for the determination of the Geiger-Muller counter response. The F1 and F8 tally applicability is briefly described. BDMG-04-02 model was validated by comparative analysis of the calculated results and experimental values of the counter responses that obtained on the UPGD-2 calibration facility. Additionally, the absolute, geometric and intrinsic registration efficiency of BDMG-04-02 was determined. The paper has been emphasized the disadvantages of using the method of direct counting of the electrons on the surface of the Geiger-Muller counter (F1).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.