Abstract

The custo screen pediatric met all criteria of the ESH-IP review 2010, modified for children from 3 to about 12years, and can be recommended for ABPM in children. What is Known: • Validation of blood pressure measuring devices is essential to provide patients with an accurate blood pressure measuring device. • The majority of devices has not been validated in children. What is New: • Prior to the present validation, study protocol adjustments of ESH-IP review 2010 for children were defined according to German Health Interview and Examination Survey for Children and Adolescents 2013 (KIGGS). • The custo screen pediatric test device met all criteria of ESH-IP revision 2010, modified for children, and can be recommended for ABPM in children aged 3 to about 12years.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.