Abstract
A total of 249 FTA ® card samples were investigated with the Identifiler ® Direct kit. PCR amplification was performed according to the manufacturer's protocol with modifications in the reaction volume (10–25 μL), the number of PCR cycles (25–27) and the number of FTA-card punches (1–2). Full concordance was observed between the results obtained with the Identifiler ® Direct kit and the washed punches typed with the Identifiler ® kit. The success rate with one 1.2 mm punch in a 15 μL PCR reaction volume and 25 PCR-cycles was 93.3%. After retyping of the failed samples, the combined success rate was 98.7%. Overall, the quality of the results was reduced when the number of punches was increased or the PCR reaction volume was decreased. The Identifiler ® Direct master mix may be used in combination with other primer sets for direct amplification of other loci, e.g. the 49 SNPs using the SNP forID multiplex assay.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Forensic Science International: Genetics Supplement Series
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.