Abstract

The objective of this study is to evaluate the accuracy of the oscillometric upper-arm device ABPMpro (SOMNOmedics) for ambulatory blood pressure measurement in the general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018) at rest and during dynamic exercise. Subjects were recruited to fulfill the age, sex, blood pressure (BP) and cuff distribution criteria of the AAMI/ESH/ISO standard using the same arm sequential BP measurement method. Three appropriate cuff sizes (18-24, 24-34 and 34-46 cm) of the tested device were used for the arm-varying circumferences. The inflation and deflation measurement modes of the ABPMpro were investigated. For the general validation study, 100 subjects were recruited and 90 were analyzed. For validation criterion (1), the mean ± SD of the differences between ABPMpro and reference BP was 0.7 ± 7.3/-0.7 ± 5.8 mmHg (systolic/diastolic) for inflation and 1.4 ± 7.7/-0.6 ± 6.1 mmHg for deflation measurements. For criterion (2), the SD of the averaged BP differences per subject was 5.98/5.10 mmHg for inflation and 6.46/5.36 mmHg for deflation measurements, thereby passing the threshold. In the ambulatory validation study ( N  = 36), the mean difference was -1.2 ± 7.9/ 2.4 ± 6.6 mmHg for inflation and -0.7 ± 7.6/3.1 ± 7.0 mmHg for deflation measurements. The ABPMpro device fulfilled the ISO 81060-2:2018 requirements in the general population and in the ambulatory setting and can therefore be recommended for clinical use.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call