Abstract

As part of a study on the core body temperature (Tb) of desert birds, we purposed to use temperature-sensitive implantable radio transmitters. Because of the difficulty in recapturing these birds, we needed to know if these electronic devices held their calibration over the duration of normal battery life (21 days). Our initial calibration at 40°C was consistently less than that of the manufacturer, although in some cases differences were small: deviations in temperature as predicted by the company's calibration and our own ranged from 0.1 to 1.7°C, We tested for drift in calibration and found that for the first 9 days of operation, most radio transmitters deviated by less than 0.5°C from the initial calibration. Differences between predicted and actual temperature became progressively larger with time, sometimes exceeding 2.0°C. If possible, we suggest that transmitters be calibrated, implanted, and then after taking data on Tb of the free-living bird, the transmitters should be recovered and recalibrate...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.