Abstract

PET-SORTEO is a Monte Carlo-based simulator that enables the fast generation of realistic PET data for the geometry of the ECAT EXACT HR+ scanner. In order to address the increasing need for simulation models for animal PET imaging, our aim is to adapt and configure this simulation tool for small animal PET scanners, especially for the three widely distributed systems manufactured by CTI-Concorde Microsystems (R4, P4 and Focus). Our goal is to propose a simulation model that will allow producing realistic rodent images in order to evaluate and optimize acquisition and reconstruction protocols. The first step presented in this study was to validate SORTEO against the geometry of the R4 system and to demonstrate the feasibility of simulating realistic activity distributions in mice in a short computation time. This validation of the simulation model was carried out against actual measurements performed on the R4 scanner at the Montreal Neurological Institute

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.