Abstract

To perform a clinical validation for a semi-automatic arm-type device UA-704 (A&D Company, Ltd., Tokyo, Japan) according to the British Hypertension Society protocol. Eighty-five study participants (37 men) were included. The mean age was 50.1+/-17.0 years, systolic blood pressure range was 79-212 mmHg and diastolic blood pressure range was 43-118 mmHg. For each participant, three readings of the UA-704 were compared with simultaneous auscultatory measurements by two trained independent observers. The observers used a mercury calibrated sphygmomanometer and a dual stethoscope. The results were graded according to the 1993 British Hypertension Society protocol. The average difference between mercury sphygmomanometer and UA-704 readings was -1.85+/-4.26 mmHg (mean+/-SD) for systolic blood pressure and -1.44+/-3.97 mmHg for diastolic blood pressure. The proportions of values agreeing to within 5, 10 and 15 mmHg were 79.2%, 96.5% and 99.6% for systolic blood pressure and 86.7%, 96.9% and 99.6% for diastolic blood pressure for the observers and device (A/A grade for British Hypertension Society). For an adult population, the UA-704 device for self-measurement of blood pressure achieved a British Hypertension Society grade A/A and therefore can be recommended for home blood pressure monitoring.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.