Abstract

The objective of this study is to validate the US Food and Drug Administration (FDA) rea-time polymerase chain reaction (qPCR) assay, the Neogen Amplified Nucleic Single Temperature Reaction (ANSR) assay, and the Vitek ImmunoDiagnostic Assay System (VIDAS) SLM procedure against the FDA cultural procedure for Salmonella detection in green chile pepper. Green chile was artificially contaminated with Salmonella according to the FDA guidelines (FDA. Guidelines for the Validation of Microbiological Methods for the FDA Foods Program, 3rd Edition. 2019. www.fda.gov/media/83812/download?attachment (17 March 2024, date last accessed)) at a fractional recovery level (where 50%-25% tests positive and at a level+1 log greater for each organism tested). Enriched samples were tested directly by the ANSR Salmonella test and by qPCR, and were subcultured into Rappaport-Vassiliadis and tetrathionate brilliant green broth for cultural detection and qPCR. For the VIDAS-SLM assay, the selective enrichments were further cultured in M broth before testing. Presumptive salmonellae were confirmed with biochemical tests, serology, and qPCR. All three rapid assays were compared favorably with the FDA-BAM (Bacteriological Analytical Manual) method. No significant differences at P<.05 were found between the procedures using McNemar's χ2 test. The three procedures were found to be rapid and reliable alternatives to cultural detection of Salmonella enterica in green chile.