Abstract
Background Accurate measurement of neonatal cardiac function remains a key goal to improve understanding of circulatory pathophysiology in prematurity. Cardiac MRI has been shown to be viable in this population, but robust validation is crucial. Our aim was to assess the accuracy of cardiac functional measures derived from MRI ventricular stack segmentation against phase contrast measures in term and preterm neonates. Methods Cardiac MRI scans were performed on a 3T Philips Achieva scanner (Best, Netherlands) in 40 infants (gestational age 24+3-40+2weeks/ weight at scan 750-3120g). Left ventricular stacks were segmented for epicardial and endocardial contours at end systole and end diastole using Segment software (http://segment.heiberg.se) with manual correction of segmentation where necessary. Correction for long axis motion was performed using the mean of three measures of atrio-ventricular valve descent from a 4 chamber view. Right ventricular stacks were segmented for endocardial contours on 20 infants (gestational age 24+3-40+2weeks/ weight at scan 7903120g) using the same technique. Cardiac output (CO) was measured by semi-automatic segmentation (View Forum,Philips) of phase contrast(PC) images of the left ventricular outflow tract. The two techniques were compared using a Bland Altman plot. Infants with known patent ductus arteriosus were excluded. Results Bland Altman analysis of left ventricular output by stack and PC showed a mean bias of 14.2 ml/min, with limits of agreement of -54.8 to 83.2 ml/min,, corresponding to a repeatability index (RI, limits of agreement/mean of population) of 16.6%. Bland Altman analysis of right ventricular output by stack and left ventricular output by PC showed a mean bias of 39.7 ml/min, with limits of agreement of -36.9 to 116.5 ml/min corresponding to a repeatability index (RI) of 17.3%. For comparison Bland Altman of neonatal heart rate during by stack and PC imaging showed a mean bias of -2.12 beats/min, with limits of agreement of -19.21 to 14.96 beats/min, corresponding to a repeatability index (RI) of 11.92%. Conclusions Given the wide variability of heart rates and potential for motion in preterm neonatal scanning, our findings suggest stack segmentation at our improved resolution can give relatively accurate measures of preterm cardiac function.
Highlights
Accurate measurement of neonatal cardiac function remains a key goal to improve understanding of circulatory pathophysiology in prematurity
Our aim was to assess the accuracy of cardiac functional measures derived from MRI ventricular stack segmentation against phase contrast measures in term and preterm neonates
Bland Altman analysis of left ventricular output by stack and PC showed a mean bias of 14.2 ml/min, with limits of agreement of -54.8 to 83.2 ml/min, corresponding to a repeatability index (RI, limits of agreement/mean of population) of 16.6%
Summary
Validation of MRI quantification of biventricular cardiac function in term and preterm neonates. Anna Finnemore*, Anthony N Price, Giuliana Durighel, Kathryn Broadhouse, David J Cox, Alan Groves. From 16th Annual SCMR Scientific Sessions San Francisco, CA, USA. From 16th Annual SCMR Scientific Sessions San Francisco, CA, USA. 31 January - 3 February 2013
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.