Abstract

ICON is the new cold neutron imaging facility at the neutron spallation source SINQ. The ICON facility is placed at beam-line S52 with direct view to the cold liquid D 2 moderator. The beam-line includes a 4.4 m long collimation section followed by a 11 m long flight path to the imaging system. The essential part of the collimation section is composed of six revolving drums and a variable aperture wheel. Depending on the investigated object, different apertures are used. Measurements have shown that each setup has a different spatial neutron flux distribution and specific beam profiles. Measured beam profiles have been used to validate results of simulations coupling the Monte-Carlo program MCNPX with the neutron ray-tracing program McStas. In a first step, MCNPX was used to calculate neutron spectra closed to the SINQ target, at the entrance of the collimation section. These results served as an input for McStas where the beam-line itself was simulated. In the present paper, experimental and theoretical results will be compared and discussed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.