Abstract
We characterize the capabilities and limitations of the Living Image Software 3D Analysis package (Xenogen, Alameda, California) in the reconstruction of calibrated light sources. Sources shallower than the mean free path of light propagation suffered reconstruction inaccuracy. For sources deeper than the mean free path, the average error in depth and intensity reconstruction was less than 4% and 12%, respectively, for homogeneous tissue. The reconstruction of luminescent beads implanted within an optically heterogeneous mouse abdomen proved less accurate. The ability to distinguish multiple sources decreased with increasing source depth. A number of factors influence the accuracy of light source reconstruction.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.