Abstract

Nondestructive digital processing methods such as lab color mode (available in Adobe Photoshop) are emerging as alternative methods for forensic document examiners to use when attempting to differentiate writing instrument inks. Although these techniques appear to be viable, little data currently exists regarding the known or potential error rates associated with these techniques. Without adequate data, the validity and reliability of these techniques, including lab color, can not be established. In an attempt to begin to address these issues, 44 black ballpoint ink pens were obtained and used to create 990 pen-pair samples for analysis using established lab color mode techniques. No erroneous findings of "different" were reported following the examination of the known pen-pair combinations in which the same pen was used to create the samples (n = 44). Of the remaining 946 samples, 737 pen-pair samples were differentiated using the lab color mode method, while 209 samples were unable to be differentiated and were recorded as either being "similar" (n = 153) or "unsure" (n = 56). Comparison of the lab color mode results with the results obtained through additional testing using traditional infrared reflectance and infrared luminescence test methods showed that lab color differentiated 102 pen-pair samples (11%; 102/946) that were not differentiated using a VSC-4C.