Abstract

Newborn screening is designed for presymptomatic identification of serious conditions with effective early interventions. Clinical laboratories must perform prospective pilot studies to ensure that the analytical performance and workflow for a given screening test are appropriate. We assessed the potential to screen newborns for fragile X syndrome, a monogenic neurodevelopmental disorder, by establishing a customized, high-throughput PCR and analysis software system designed to detect fragile X mental retardation 1 gene repeat expansions from dried blood spots (DBSs). Assay precision, accuracy, sensitivity, and specificity were characterized across the categorical range of repeat expansions. The assay consistently resolved genotypes within three CGG repeats of reference values up to at least 137 repeats and within six repeats for larger expansions up to 200 repeats. Accuracy testing results were concordant with reference results. Full and premutation alleles were detected from subnanogram DNA inputs eluted from DBSs and from mixtures with down to 1% relative abundance of the respective expansion. Analysis of 963 deidentified newborn DBS samples identified 957 normal and 6 premutation specimens, consistent with previously published prevalence estimates. These studies demonstrate thatthe assay system can support high-throughput newborn screening programs.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call