Abstract

ABSTRACT Introduction To speed up antimicrobial susceptibility testing (AST), the European Committee on Antimicrobial Susceptibility Testing (EUCAST) proposed rapid AST (RAST), a disk diffusion method to be read after 4, 6 and 8 hours of incubation. We investigated the feasibility of implementation of RAST in a non-automated lab setting. Materials & methods To this end, reference strains as well as a variety of clinical and resistant strains were used to spike sterile hemocultures (BioMérieux BACT/ALERT 3D® and Becton Dickinson BACTEC FX® systems), followed by RAST in comparison to classical long-incubation AST. Results & conclusion Our results with reference strains show that reading RAST after 4 hours is frequently too soon to obtain clinical results, and that Streptococcus pneumoniae reference strain did yield readable inhibition zones in RAST when harvested from BioMérieux BACT/ALERT 3D® bottle cultures. In a wider panel of strains, Gram positives RAST results were very similar to standard AST, while with Gram negative species errors were more frequently observed, limiting clinical implementation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.