Abstract

HIV viral load (VL) and donor screening assays experience variation and require quaity assurance (QA). NRL sought to confirm a dried tube sample format (HIVDTS) sample type for use in quality control (QC) programs for HIV molecular testing.50 μL of HIV supernatant at 1 × 105 copies per millilitre (copies/mL)) was dried for 48 hours at room temperature. Post-production and shipped integrity studies were undertaken. Dried HIVDTS was reconstituted in PBS buffer and tested in HIV VL (six participants) or blood screening assays (four participants). Results were entered into NRL’s QC monitoring software (EDCNet™) for analysis.The mean of 224 VL results when HIVDTS QCs were tested in Biocentric HIV GENERIC Charge Virale assay was 4.54 log10 copies/mL, with the percentage coefficient of variation (CV%) ranging from 1.75 to 13.20%. The mean Ct value for HIVDTS QCs tested on Roche Cobas MPX assay results was 28.71 (range 28.33 to 29.14), with CV% ranging from 1.56 to 3.98%.The study confirms HIVDTS QCs can effectively monitor the performance of HIV molecular testing and offers a cheaper alternative to commercial QC samples that require cold-chain shipping on dry ice and UN3373 conditions.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.