Abstract

This paper is concern with a reverse phase high performance liquid chromatography (RP-HPLC) bio-analytical method development and validation for Prasugrel in human plasma using photo diode array detector (PDA detector). The HPLC separation was carried out in an isocratic mode on an X-Terra C18 column (4.6 x 150 mm; 5 μm) with a mobile phase consisting of potassium dihydrogen phosphate [pH 3.0] and acetonitrile in the ratio of 30:70 v/v at a flow rate of 1.0 mL/min. The run time was maintained for 5 mins and the detection was monitored at 210 nm. The percentage recovery was found 99.61-100.06 in human plasma. This reveals that the method is quite accurate. The linearity was found 15-40 μg/mL in human plasma. The inter-day and intra-day precision in plasma was found within the limits. The lower limit of quantification (LLOQ) obtained by the proposed method was 0.05 μg/mL. The percentage relative standard deviation (%RSD) obtained for the drug spiked in plasma for stability studies were less than 2 %.Kathmandu University Journal of Science, Engineering and TechnologyVol. 13, No. 1, 2017, Page: 65-75

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.