Abstract

The coupling of capillary electrophoresis (CE) to mass spectrometry (MS) with an electrospray ionization (ESI) technique is generally performed for qualitative applications, and only few quantitative results have been reported. This paper investigates the validation of a CE-ESI-MS method for the analysis of a pharmaceutical formulation containing lidocaine. Some important ESI criteria are discussed including sheath-liquid composition, nebulizing gas pressure and position of the CE capillary outlet. After optimization of these parameters, an intermediate precision of about 5% was achieved. The latter, as well as efficiency and resolution, were compared to those achieved with UV detection. Besides, a multiple injection procedure was developed to reduce analysis time per sample and was successfully applied to both UV and MS detectors. The validation results achieved by multiple injections were identical to those obtained with classical injection, but afforded a gain of time by a factor of 2.5.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.