Abstract

The method of determining calcium (Ca) in dolomite fertilizer has been validated using atomic absorption spectrophotometer (AAS). Determining calcium (Ca) in dolomite fertilizer refers to the Indonesian National Standard (SNI) 02-2804-200. Based on the results of the calcium (Ca) test on dolomite fertilizer, the correlation coefficient (r) was 0.9949, and the % relative standard deviation (RSD) was 0.6472%, less than 2/3 CV Horwitz of 1.5963. The accuracy value is 100.30%, the precision value is 0.6471%, the detection limit for the LOD method is 0.2817, the LOQ is 0.9390, and the instrument detection limit is LOD 0.1159 and LOQ 0.3864. This study shows that the determination of calcium (Ca) in dolomite fertilizer using the SNI method using the Atomic Absorption Spectrophotometer (AAS) can be used with valid results.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.