Abstract
Biologically active amines include the so called biogenic amines, such as histamine, tyramine and cadaverine, and polyamines such as spermidine and spermine. Ultra high pressure liquid chromatography (UHPLC) is a new generation of separation techniques that takes full advantage of chromatographic principles to increase speed flow which drastically reduce analysis time. The aim of the present work was to validate a rapid method of UHPLC to detect the presence of biogenic amines and polyamines in food. Different food matrixes (wine, fish, cheese, and dry fermented sausage) were used in order to test the versatility of the method. The UHPLC method described in this article has been demonstrated as a reliable procedure to determine 12 biogenic amines and polyamines in less than 7 min of chromatographic elution. The method provides a satisfactory linearity and chromatographic sensitivity with a detection limit lower than 0.2 mg/L and a determination limit falling below 0.3 mg/L for all amines. The precision, in terms of relative standard deviation, was lower than 5% and the accuracy, as mean recovery, was between 93% and 98%, depending on the food matrix.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.