A liquid chromatographic procedure using ion interaction reagent and evaporative light scattering detection was developed and validated for the direct quantification of 8 underivatized amino acids (threonine, lysine, valine, methionine, isoleucine, leucine, phenylalanine and tryptophan) present simultaneously in a commercial preparation. The chromatographic separation was achieved on a Purospher RP-18e column with an acetonitrile gradient and using heptafluorobutyric acid as volatile ion interaction reagent. Acceptable levels of linearity, specificity, precision, accuracy, robustness and limits of detection were achieved during the validation of the method. Detection limits varied from 1 mg.L−1 to 2 mg.L−1 depending on the amino acid being analyzed. The results of this method agree to within 98.6%–102.5% when compared to the manufacturer’s certificate of analysis obtained with the standard amino acid autoanalyser method. This approach yields a simple, universal method that is well suited for amino acid analysis, when a sufficient quantity is available. Only basic LC instrumentation with an ELSD is necessary for this procedure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.