Abstract

The aim of this investigation was to evaluate the agreement between a manual and an automatic technique in assessing levator hiatus area (LHA) during pregnancy from three-dimensional (3D) pelvic floor volumes obtained by trans-perineal ultrasound (TPUS). 3D volumes were acquired during rest, maximum pelvic floor contraction and Valsalva maneuver from 66 pregnant women. Manual selection of LHA and automatic software (Smart Pelvic™) were applied on TPUS volume starting from a C-plane view. To evaluate intra- and inter-observer variability measurements of LHA were performed twice by the same operator and once by a second sonographer. Reference hiatal contours obtained manually by the first operator were compared with the automated ones. Reproducibility was evaluated by intraclass correlation coefficients (ICC) and Bland-Altman plots. LHA measurement, using automatic software, achieved excellent intra-observer and inter-observer reproducibility in pregnant women both at rest and after dynamic analysis (ICC>0.9). Further, an excellent agreement resulted between manual selection of the LHA and automatic imaging (ICC>0.9). The average time taken to obtain LHA manually was significantly longer when compared to the automatic analysis (p≤0.0001). Smart pelvic software resulted from a reliable method for automatically measuring the LHA, showing high reproducibility and accuracy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.