Abstract

The performance characteristics and diagnostic accuracy of a new rapid automated quantitative immunoturbidimetric D-dimer assay, Auto-Dimer (Biopool, Umeå, Sweden) were evaluated in a population of 135 outpatients with suspected deep-vein thrombosis of a lower limb. Enzyme-linked immunosorbent assay was used as the reference method. The Auto-Dimer assay showed good reproducibility. The correlation between Auto-Dimer and enzyme-linked immunosorbent assay was high (r = 0.91, p < 0.05) and agreement in classifying patients above or below the cut-off was good (kappa coefficient 0.74, 95% CI 0.62-0.86). At a cut-off of 340 microg/l the sensitivity and specificity of Auto-Dimer were high (100% and 61%, respectively). These data show that Auto-Dimer is a reliable screening test for exclusion of deep-vein thrombosis. The assay could be included in prospective patient management studies in order to obtain further information on its clinical usefulness.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.