Abstract

It has been implement a noninvasive technique to detect and delineate the focus of electric discharge in patients with mono-focal epilepsy. For the detection of these sources it has used electroencephalogram (EEG) with 128 electrodes cap. With EEG data and electrodes position, it is possible locate this focus on MR volumes. The technique locates the electrodes on CT volumes using image processing algorithms to obtain descriptors of electrodes, as centroid, which determines its position in space. Finally these points are transformed into the coordinate space of MR through a registration for a better understanding by the physician. Due to the medical implications of this technique is of utmost importance to validate the results of the detection of electrodes coordinates. For that, this paper present a comparison between the actual values measured physically (measures including electrode size and spatial location) and the values obtained in the processing of CT and MR images.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.