Abstract

We have used an implanted telemetry system to continuously monitor blood pressure (BP) in three dogs for durations ranging from 28 to 75 wk after implantation. Measurements of BP obtained by telemetry were compared every 3-12 wk, with measurements of BP recorded with a manometer-tipped catheter that was inserted into a femoral artery. Over a wide range of both physiological and pharmacologically manipulated pressures (40-200 mmHg), the values of BP obtained by the two methods were highly correlated (all r > 0.966; all P < or = 0.0001). However, the mean differences between the values obtained by the two systems were different from zero (range +29.6 to -1.5 mmHg; P < or = 0.0001), indicating an offset in the BP values recorded from the implanted system. Furthermore, this offset was dependent on the absolute level of the BP. The findings indicate that, for a period of at least 28 wk and up to 75 wk after implantation, the telemetry system accurately measures acute changes in BP and can reliably measure absolute BP provided that the system is properly validated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.