Abstract

Iodine is an essential part of the thyroid hormones thyroxine and triiodothyronine. Therefore, it is essential to monitor iodine supply in a population. The biochemical marker for assessing and controlling iodine is urinary iodine concentration (UIC). This cross-sectional study included 180 pregnant women and 308 women of reproductive age. Urine specimens from 185 of the 488 volunteers were used. The urine specimens were measured using 2 methods: (1) ammonium persulfate digestion (APD), followed by the Sandell-Kolthoff (S-K) reaction modified on microplate for spectrophotometric detection; and (2) the reference method, inductively coupled plasma mass spectrometry (ICP-MS). The regression equation between the methods was ICP-MS method = 1.137*(APD S-K)-5.57. A Passing-Bablok regression showed no deviation from linearity (P = .17). A Bland-Altman plot showed a negative mean bias of -2.7%. The APD S-K reaction modified on microplate for spectrophotometric detection of UIC can be implemented into routine work. Its results are comparable to those of laboratories worldwide and to ICP-MS.