Abstract

Our objective was to validate a point-of-care handheld blood total calcium analyzer (Ca meter, CM; TD-5220 Vet Ca2+, TaiDoc, New Taipei, Taiwan) to estimate circulating Ca concentrations in postpartum dairy cows. Whole blood was collected from 251 multiparous cows between 1 and 4 d in milk from 2 commercial dairy herds in Ontario, Canada. Blood total calcium concentration (tCa) was analyzed in whole blood, fresh plasma, and thawed plasma, and compared with tCa results from thawed serum analyzed in a diagnostic laboratory (using a Cobas Calcium Gen 2 kit, Roche Diagnostics, Indianapolis, IN) as the reference test (RT). Lin's concordance correlation coefficient (βrho;) and Bland-Altman (B-A) plots were assessed to evaluate the agreement between the RT and CM results in each type of sample. Receiver operating characteristic curve analyses were used to describe the accuracy of each test against the categorized RT results (at a cut-point of ≤2.14 mmol/L). Samples where the meter gave a nonquantitative result (“high” or “low”; thawed plasma: 3/247; fresh plasma: 6/100; and whole blood: 20/98) were not included in the βrho; and B-A analyses. Lin's correlation coefficients demonstrated poor agreement between tests (thawed plasma: βrho; = 0.16; fresh plasma: βrho; = 0.21; and whole blood: βrho; = 0.23). Fresh plasma (using a cut-point of 2.55 mmol/L as measured on the CM) had the greatest diagnostic sensitivity (72%), specificity (86%), and accuracy (77%) for determining subclinical hypocalcemia, but that would still misclassify 23% of samples. In addition to substantial variability, the B-A plots revealed bias with changing concentrations of calcium. Because of low sensitivity on whole blood (58%) or thawed plasma (56%), measurement with the CM is not recommended on these types of samples. This rapid and low-cost meter was not sufficiently accurate to quantify blood Ca concentration, but when used with fresh plasma it might be useful as a screening tool for subclinical hypocalcemia.

Highlights

  • Blood total calcium concentration was analyzed in whole blood, fresh plasma, and thawed plasma, and compared with tCa results from thawed serum analyzed in a diagnostic laboratory as the reference test (RT)

  • Lin’s concordance correlation coefficient (ρ) and Bland-Altman (B-A) plots were assessed to evaluate the agreement between the RT and CM results in each type of sample

  • Receiver operating characteristic curve analyses were used to describe the accuracy of each test against the categorized RT results

Read more

Summary

Introduction

Abstract: Our objective was to validate a point-of-care handheld blood total calcium analyzer (Ca meter, CM; TD-5220 Vet Ca2+, TaiDoc, New Taipei, Taiwan) to estimate circulating Ca concentrations in postpartum dairy cows. Samples where the meter gave a nonquantitative result (“high” or “low”; thawed plasma: 3/247; fresh plasma: 6/100; and whole blood: 20/98) were not included in the ρ and B-A analyses.

Objectives
Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.