Abstract

In this paper the experimental setup of a new apparatus able to provide the critical properties of pure components and mixtures using either a dynamic or a static method is described. Critical temperatures (Tc) and critical pressures (Pc) of pure components (n-pentane, n-heptane and n-decane) and critical loci of three binary mixtures (n-pentane + n-heptane, n-pentane + n-decane and n-heptane + n-decane) are investigated using the dynamic method. For the system n-pentane + n-heptane, the results obtained with the dynamic method are compared with those obtained with the static method. The critical points are visually determined by observing the critical opalescence and the simultaneous disappearance and reappearance of the meniscus in the middle of the view cell which withstands operations up to 673 K and 20 MPa. The experimental critical data are compared with success to literature database and with their prediction from the PPR78 thermodynamic model.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.