Abstract
Pulmonary vein isolation (PVI) is the cornerstone of ablation for atrial fibrillation. Confirmation of PVI can be challenging due to the presence of far-field electrograms (EGMs) and sometimes requires additional pacing manoeuvres or mapping. This prospective multicentre study assessed the agreement between a previously trained automated algorithm designed to determine vein isolation status with expert opinion in a real-world clinical setting. Consecutive patients scheduled for PVI were recruited at four centres. The ECGenius electrophysiology (EP) recording system (CathVision ApS, Copenhagen, Denmark) was connected in parallel with the existing system in the laboratory. Electrograms from a circular mapping catheter were annotated during sinus rhythm at baseline pre-ablation, time of isolation, and post-ablation. The ground truth for isolation status was based on operator opinion. The algorithm was applied to the collected PV signals off-line and compared with expert opinion. The primary endpoint was a sensitivity and specificity exceeding 80%. Overall, 498 EGMs (248 at baseline and 250 at PVI) with 5473 individual PV beats from 89 patients (32 females, 62 ± 12 years) were analysed. The algorithm performance reached an area under the curve (AUC) of 92% and met the primary study endpoint with a sensitivity and specificity of 86 and 87%, respectively (P = 0.005; P = 0.004). The algorithm had an accuracy rate of 87% in classifying the time of isolation. This study validated an automated algorithm using machine learning to assess the isolation status of pulmonary veins in patients undergoing PVI with different ablation modalities. The algorithm reached an AUC of 92%, with both sensitivity and specificity exceeding the primary study endpoints.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.