Abstract

Diuron is one of the most widely herbicide used worldwide, which can undergo degradation producing three primary metabolites: 3,4-dichlorophenylurea, 3-(3,4-dichlorophenyl)-1-methylurea, and 3,4-dichloroaniline. Since the persistence of diuron and its by-products in ecosystems involves risk of toxicity to environment and human health, a reliable quantitative method for simultaneous monitoring of these compounds is required. Hence, a simple method without preconcentration step was validated for quantitation of diuron and its main metabolites by high performance liquid chromatography with ultraviolet detection. Separation was achieved in less than 11 minutes using a C18 column, mobile phase composed of acetonitrile and water (45:55 v/v) at 0.86 mL min-1 and detection at 254 nm. The validated method using solid-liquid extraction followed by an isocratic chromatographic elution proved to be specific, precise and linear (R2 ˃ 0.99), presenting more than 90% of recovery. The method was successfully applied to quantify diuron and their by-products in soil samples collected in a sugarcane cultivation area, focusing on the environmental control.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.