Abstract

A rapid and sensitive RP-HPLC method with UV detection for routine control of oxaliplatin in a pharmaceutical formulation (Eloxatin®) was developed. Quantitation was accomplished with the internal standard method. The procedure was validated by linearity (correlation coefficient=0.999948), accuracy, robustness and intermediate precision. Experimental design was used during validation to calculate method robustness and intermediate precision. For robustness test three factors were considered: percentage v/v of acetonitrile, flow rate and temperature; an increase in the flow rate results in a decrease of the drug found concentration, while the percentage of organic modifier and temperature have no important effect on the response. For intermediate precision measure the considered variables were: analyst, equipment and days. The RSD value (2.27%, n=24) indicated a good precision of the analytical method.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.