Abstract

The expansion and increasing availability of computed tomography (CT) imaging means that there is a greater need for the development of efficient optimisation strategies that are able to inform clinical practice, without placing a significant burden on limited departmental resources. One of the most fundamental aspects to any optimisation programme is the collection of patient dose information, which can be compared with appropriate diagnostic reference levels. This study has investigated the implementation of a large-scale audit technique, which utilises data that already exist in the radiology information system, to determine typical doses for a range of examinations on four CT scanners. This method has been validated against what is considered the 'gold standard' technique for patient dose audits, and it has been demonstrated that results equivalent to the 'standard-sized patient' can be inferred from this much larger data set. This is particularly valuable where CT optimisation is concerned as it is considered a 'high dose' technique, and hence close monitoring of patient dose is particularly important.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.