Abstract
BackgroundHLA-A2 tetramer flow cytometry, IFNγ real time RT-PCR and IFNγ ELISPOT assays are commonly used as surrogate immunological endpoints for cancer immunotherapy. While these are often used as research assays to assess patient's immunologic response, assay validation is necessary to ensure reliable and reproducible results and enable more accurate data interpretation. Here we describe a rigorous validation approach for each of these assays prior to their use for clinical sample analysis.MethodsStandard operating procedures for each assay were established. HLA-A2 (A*0201) tetramer assay specific for gp100209(210M) and MART-126–35(27L), IFNγ real time RT-PCR and ELISPOT methods were validated using tumor infiltrating lymphocyte cell lines (TIL) isolated from HLA-A2 melanoma patients. TIL cells, specific for gp100 (TIL 1520) or MART-1 (TIL 1143 and TIL1235), were used alone or spiked into cryopreserved HLA-A2 PBMC from healthy subjects. TIL/PBMC were stimulated with peptides (gp100209, gp100pool, MART-127–35, or influenza-M1 and negative control peptide HIV) to further assess assay performance characteristics for real time RT-PCR and ELISPOT methods. Validation parameters included specificity, accuracy, precision, linearity of dilution, limit of detection (LOD) and limit of quantification (LOQ). In addition, distribution was established in normal HLA-A2 PBMC samples. Reference ranges for assay controls were established.ResultsThe validation process demonstrated that the HLA-A2 tetramer, IFNγ real time RT-PCR, and IFNγ ELISPOT were highly specific for each antigen, with minimal cross-reactivity between gp100 and MelanA/MART-1. The assays were sensitive; detection could be achieved at as few as 1/4545–1/6667 cells by tetramer analysis, 1/50,000 cells by real time RT-PCR, and 1/10,000–1/20,000 by ELISPOT. The assays met criteria for precision with %CV < 20% (except ELISPOT using high PBMC numbers with %CV < 25%) although flow cytometric assays and cell based functional assays are known to have high assay variability. Most importantly, assays were demonstrated to be effective for their intended use. A positive IFNγ response (by RT-PCR and ELISPOT) to gp100 was demonstrated in PBMC from 3 melanoma patients. Another patient showed a positive MART-1 response measured by all 3 validated methods.ConclusionOur results demonstrated the tetramer flow cytometry assay, IFNγ real-time RT-PCR, and INFγ ELISPOT met validation criteria. Validation approaches provide a guide for others in the field to validate these and other similar assays for assessment of patient T cell response. These methods can be applied not only to cancer vaccines but to other therapeutic proteins as part of immunogenicity and safety analyses.
Highlights
HLA allele A*0201. MART-1 (HLA-A2) tetramer flow cytometry, IFNγ real time reverse transcription (RT)-PCR and IFNγ enzyme linked immunospot assay. Gp100 (ELISPOT) assays are commonly used as surrogate immunological endpoints for cancer immunotherapy
Tetramer assay specificity is defined as tumor infiltrating lymphocyte cell lines (TIL) cells which lack binding to negative tetramer and irrelevant tetramer and show specific binding to the relevant tetramer (TIL1520 binding to gp100 and TIL1143 binding to MART-1)
To confirm that MART-1 tetramer can maximally detect all of the MART-1 specific T cells under the assay conditions used, Jurkat cells that were genetically modified to express TCR that recognizes MART-1/HLA-A2 were used and 97% of MART-1 tetramer positive cells were detected; irrelevant gp100 tetramer binding to the MART-1 Jurkat cells was minimal (0.04%), Figure 2(B)
Summary
HLA-A2 tetramer flow cytometry, IFNγ real time RT-PCR and IFNγ ELISPOT assays are commonly used as surrogate immunological endpoints for cancer immunotherapy. While these are often used as research assays to assess patient's immunologic response, assay validation is necessary to ensure reliable and reproducible results and enable more accurate data interpretation. Assay development studies (IFNγ Real Time RT-PCR and ELISPOT, HLA-A2 Tetramer analysis) and monitoring specific vaccine response in cancer patients are described by a number of investigators [2,3,4,5,6,7,8,9,10]. The validation of immunoassays was identified as one of the critical areas for improvement when using these assays to evaluate immune responses in the clinic [1]
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.