Abstract
Methadone is a synthetic opiate derivative that possesses analgesic activity. A modified fluorescence polarization immunoassay (FPIA) method and a high-pressure liquid chromatography (HPLC) method with UV detection were compared for measurement of concentrations of methadone in canine plasma following intravenous and oral methadone administration. The mean+/-SD for accuracy (deviation from actual concentration) and precision (coefficient of variation) when methadone-fortified canine plasma was evaluated with the FPIA method were 3.9+/-3.2% and 4.4+/-2.9%, respectively. The accuracy and precision of the HPLC method were 6.2+/-5.2% and 7.7+/-3.9%, respectively. The limit of quantification for the FPIA and HPLC methods were 25 and 20 ng/mL, respectively. The coefficient of determination (r) between FPIA and HPLC analysis was 0.94 when plasma from dogs dosed with methadone was evaluated. FPIA provides a rapid, sensitive, and specific measurement of methadone in canine plasma following oral and intravenous administration.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.