Abstract
An anion exchange high-performance liquid chromatography (HPLC) method for the quantification of human Reovirus type 3 particles was validated according to the performance criteria of precision, specificity, linearity of calibration and working range, limits of detection and quantification, accuracy and recovery. Samples taken at various stages of Reovirus purification were used for the validation of the method. The method was specific for Reovirus which eluted around 9.8 min without interference from any other component in the sample. Reovirus can be detected between 0.32E+12 and 2.10E12 VP/mL by the proposed method that has the correlation coefficient of linearity equal to 0.9974 and the slope of linearity equal to 5.74E−07 area units/(VP mL).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.