Abstract

For many years, fluorescence polarization immunoassay (FPIA) on the TDx analyzer has been used for determination of free phenytoin concentration. Recently Abbott Laboratories decided to discontinue the TDx analyzer and related assays on this analyzer. Free phenytoin assay is also available from Roche Diagnostics for application on the Cobas Integra analyzer (fluorescence polarization assay) but not on Cobas c510 analyzer. Free phenytoin calibrators from the Cobas Integra free phenytoin assay and the reagents from the KIMSphenytoin assay were used for the determination of free phenytoin on the Cobas c501 analyzer. The intra-run and inter-run precisions were both <7.2%. The assay was linear from 0.2 to 4 μg/ml. The free phenytoin assay on the Cobas c501 was compared with the FPIAassay on the TDx analyzer using sera from 25 patients receiving phenytoin (phenytoin concentration between 0.3 and 3.7 μg/ml). The following regression equation was observed: y = 0.9899 x + 0.0408 (r = 0.98, n = 25). In conclusion, the free phenytoin assay on the Cobas c501 analyzer is a valid alternative to free phenytoin assay on the TDx analyzer.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.