Abstract

Computer-assisted surgery is becoming more frequently used in the medical world. OrthoPilot™ Hip Suite CT-free navigation system (BBraun Aesculap, Tuttlingen, Germany) is one such computer-assisted navigation system used for total hip replacement surgery. The validity of OrthoPilot system remains to be determined independent of the manufacturer. The main aims of this study were to investigate the accuracy of OrthoPilot data while using surgical instruments and to confirm the suitability of the cup navigation algorithm. OrthoPilot was compared with the gold standard of a VICON movement analysis system. An aluminium pelvic phantom was machined with high accuracy to perform the experiments. Data were captured simultaneously from both OrthoPilot and VICON systems and acetabular angles were compared. Both systems produced comparable results for the distance between anatomical landmarks and acetabular angles. It can be concluded that data from the OrthoPilot system, if used correctly, are accurate enough for orthopaedic applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.